1100 Series DAD and MWD Reference Manual 109
Repairing the Detector 4
Note:
If you want to replace the gasket only, continue with
step 8.
Do not mix the gasket #6 and # 7. They are different
for standard and semi-mirco flow cell.
2
Use a tooth pick to remove the quartz window from the
window assembly.
3 If the washers fall out of the window assembly, they must be inserted in the correct order with the Teflon® ring to prevent
any leaks from the flow cell window.
Orientation of Flow Cell Parts
1
2
3
4
5
6
7
1 window screw
2 spring washers
3 compression washer
4 window holder
5 quartz window
6 gasket
7 gasket
8 window screw
contains items 2, 3, 4
and 5
9 flow cell
10 inlet capillary
11 outlet capillary
8
9
10
11
NOTE
Gaskets # 6 and #7 have
different hole diameters
Przeglądanie stron 108