1100 Series DAD and MWD Reference Manual 161
Repairing the Detector 4
12 Before installing the filter motor, press the filter lever in
from the other side.
13 While counter-holding the filter lever, insert the filter
motor onto the filter shaft until it clicks into the final
position.
14 Slide in the cell support assembly until it is stopped by the
holmium filter. Leave about 0.2-mm clearance for
movement of the filter lever between both the lens
assembly and the filter lever, and fix the setscrew.
Next steps:
• Check that the filter lever is moveable.
• Reinstall the optical unit, see “Installing the Optical Unit
(G1315A/G1365A)” on page 171 or “Removing the
Optical Unit (G1315B/G1365B)” on page 156.
• Reinstall the flow cell, see “Exchanging a Flow Cell” on
page 105.
• Reinstall the front cover, top cover and top foam section,
see “Exchanging the Detector Main Board” on page 143.
• Replace detector into the stack.
• Reconnect the cables and turn on the detector.