1100 Series DAD and MWD Reference Manual 121
Repairing the Detector 4
5 Fix the new heat exchanger to the clamp unit and the heat
exchanger to the cell body.
6 Fix the inlet capillary to the flow cell body handtight first.
Then do a 1/4 turn with a 4-mm wrench.
Next steps:
Reconnect the capillaries, see “Exchanging a Flow Cell” on page 105.
Perform a leak test.
Insert the flow cell.
Replace the front cover.
Perform a wavelength calibration, see “Wavelength Recalibration” on page 84 or holmium test, see “Holmium Oxide
Te s t ” on page 86 to check the correct positioning of the flow cell.
Inlet capillary