112 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Repairing the High Pressure Flow Cell
When required If the flow cell needs repair due to leaks or contaminations (reduced light
throughput)
Tools required Two 1/4 inch wrenches for capillary connections
hexagonal key 4 mm
Too t h pi c k s
Parts required For parts see “High Pressure Flow Cell” on page 190
Preparations for this procedure:
Turn off the flow.
Remove the front cover.
Remove the flow cell, see “Exchanging a Flow Cell” on
page 105.
1 Remove the window assembly (1) and the gasket (2) from
the cell body.
2
1
1
2