1100 Series DAD and MWD Reference Manual 173
Repairing the Detector 4
Installing the Optical Unit (G1315B/G1365B)
When required When repairs have been completed
Tools required Screwdriver POZI 1 PT3
NOTE
Use this procedure for the the G1315B/G1365B only. For G1315A/G1365A detectors refer
to “Installing the Optical Unit (G1315A/G1365A)” on page 171.
CAUTION
The fan must be installed in the correct orientation to ensure correct cooling and
operation of the detector, see “Replacing Fan Parts (G1315B/G1365B)” on page 150.
Preparations for this procedure:
The power supply, the fan and the processor board are
already installed.
1 Check that the shock absorbers and the leak sensor cable
is placed correctly.