154 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Removing the Optical Unit (G1315A/G1365A)
1 Turn off the lamp.
2 Switch off detector, and disconnect power cable.
3 Remove detector from stack and place it on working bench.
4 Remove front cover, top cover and top foam section, see “Removing the Top
Cover and Foam” on page 140.
5 Remove both lamps from the optical unit, see “Exchanging a Lamp” on
page 102 (G1315A/G1365A only!).
6 Remove flow cell, see “Exchanging a Flow Cell” on page 105.
7 Remove flow cell cover, see “Cleaning or Exchanging the Holmium Oxide
Filteron page 127 (G1315A/G1365A only!).
NOTE
Use this procedure for the the G1315A/G1365A only. For G1315B/G1365B detectors refer
to “Removing the Optical Unit (G1315B/G1365B)” on page 156.
When required For all repairs inside the optical unit
Tools required Screwdriver POZI 1 PT3
Parts required Optical unit G1315-69002 (exchange assembly) or
Individual parts depending on the following procedures
NOTE
Use this procedure for the G1315A/G1365A only.