Agilent Technologies 1100 Series Instrukcja Obsługi Strona 87

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 318
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 86
1100 Series DAD and MWD Reference Manual 87
Troubleshooting and Test Functions 3
Test Failed
Probable Causes
Absorbing solvent or air bubble in flow cell.
Incorrect calibration
Dirty or contaminated flow cell.
Dirty or contaminated optical components (achromat, windows).
Old or non-Agilent lamp.
Suggested Actions
Ensure the flow cell is filled with water.
Recalibrate (see “Wavelength Recalibration” on page 84) and repeat the
test.
Run the cell test (see “Cell Test” on page 90). If the test fails, exchange the
flow cell windows.
Clean optical components with alcohol and lint-free cloth.
Exchange the UV lamp.
Figure 22 Holmium Oxide Test Results
Przeglądanie stron 86
1 2 ... 82 83 84 85 86 87 88 89 90 91 92 ... 317 318

Komentarze do niniejszej Instrukcji

Brak uwag