114 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Replacing Capillaries on a Standard Flow Cell
When required If the capillary is blocked
Tools required Two 1/4 inch wrenches for capillary connections
Wrench 4 mm for capillary connections
Screwdriver Pozi 1 PT3
Parts required For parts see “Standard Flow Cell” on page 186
Preparations for this procedure:
Turn off the flow.
Remove the front cover.
Remove the flow cell, see “Exchanging a Flow Cell” on
page 105.
1 Identify the inlet and outlet capillaries.
Outlet capillary
Inlet
capillary