Agilent Technologies 1100 Series Instrukcja Obsługi Strona 39

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 318
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 38
1100 Series DAD and MWD Reference Manual 39
How to optmize the Detector 2
Slit Width
The detector has a variable slit at the entrance of the spectrograph. This is an
effective tool to adapt the detector to changing demand of different analytical
problems.
A narrow slit provides spectral resolution for analytes with very fine
structures in the absorbance spectrum. An example of such a spectrum is
benzene. The five main absorbance bands (fingers) are only 2.5 nm wide and
just 6 nm apart from each other.
Figure 12 Gradient Analysis of PTH-Amino Acids (1 pmol each), with and without Refer-
ence
1pmol each
Wavelength
267 nm
Reference 380 nm
Wavelength
267 nm
No reference
Time (min)
PTH-ASN
PTH-ARG
PTH-ALA
PTH-PRO
PTH-PHE
Grad.: 0.02 m KH
2
PO
4
/ACN from 12% ACN to 45% ACN in 12 min
Przeglądanie stron 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 317 318

Komentarze do niniejszej Instrukcji

Brak uwag