Optimizing the mask testing accuracy
The testing accuracy can be divided into two parts: vertical accuracy and
horizontal accuracy.
Vertical accuracy
The HP 54600-series oscilloscopes have fully calibrated vertical verniers. To
achieve the highest accuracy, the vertical sensitive should be placed in the
vernier mode and adjusted so that the input signal occupies the entire screen
(full scale).
For testing a DC signal, use the fully calibrated position control to get rid of
the dc component. With a x10 probe, a maximum of 20 volts of calibrated
offset can be used even at the most sensitive setting.
Horizontal accuracy
The HP 54600-series of oscilloscopes have fully calibrated time base verniers.
For testing time related parameters such as frequency and rise time, it is
recommended that the portion of the signal that contains the parameter
occupies approximately the whole screen width. Figure 23 shows how to set
up the oscilloscope to test rise time.