To prevent the operator from changing the setup
Once the setups for each measurement step are properly set up, you may not
want the operator to change anything in the setup, especially when the mask
is related to a particular sensitivity and position setting. However, in some
other situations, you may want the operator to make fine adjustments to fit
the signal inside the testing mask. This is especially true with a mask that
tests the shape of the signal but not the absolute amplitude. The Test
Automation Module allows you to choose three different levels of operator
permission.
•
All Allows the operator to change all of the setups in this step. All the
knobs and keys can be operated by the operator.
•
Adjust Allows the operator to operate the following knobs: Delay,
Time/Div, Volts/div, Position, and the softkeys. All of the other menu keys
are disabled. This allows the operator to fine adjust the fit of the signal
inside the mask. The Volt/Div and Time/Div verniers can be turned on in
the setup for fine adjustments.
•
None The operator is not allowed to change the setup. All of the knobs
and keys except the softkeys and the knob closest to the Cursors key are
disabled.
1 Press Print/Utility twice.
2 Press the Sequence Admin softkey.
3 Press the Define Sequence softkey.
4 Press the Step softkey, or turn the knob closest to the Cursors key,
to select the step number that you want to modify.
5 Press the Define Step softkey.
6 Press the Test Modes softkey.
7 Press the Permit softkey until the desired access ability is shown.
Operating the Test Automation Module
To prevent the operator from changing the setup
49