To set up the mask template testing for "Fail when In"
testing
Normally, the mask template test signals pass when the signal lies inside the
maximum and minimum limit. However, in certain cases, such as eye pattern
testing, you may want to set up a region in which the signal fails the test
when it lies inside this region. The Test Automation Module allows you to
select the failure condition to be inside or outside the maximum and
minimum lines.
1 Press Print/Utility twice.
2 Press the Sequence Admin softkey.
3 Press the Define Sequence softkey.
4 Press the Step softkey, or turn the knob closest to the Cursors key,
to select the step number for the testing mask number.
5 Press the Define Step softkey.
6 Press the Test Modes softkey.
7 Toggle the Fail when softkey to select In.
Operating the Test Automation Module
To set up the mask template testing for "Fail when In" testing