Step Branching operating hint
In the case where no testing mask is specified in a particular step, the module
always treats the test result as PASS and branches to the step specified in the
pass condition.
The "Same" setting in the branching destination branches back to the same
step. This allows you to repeat a step until a certain test result is found. For
example, if the branching destination of Fail Min and Fail Max are both set to
"Same," the sequencing does not branch to another step unless the signal
passes the test or the operator aborts the sequencing by pressing the Reset or
Stop Sequencing softkey.
If the signal fails both minimum and maximum limit lines, the branching path
corresponding to Fail Min is chosen.
When the fail region is defined as "In," the branching path corresponding to Fail
Min is selected for a failure condition.
Operating the Test Automation Module
To create a branching sequence
48