2 Press Print/Utility .
If the Previous Menu softkey appears on the rightmost softkey,
press
Print/Utility again.
When the module is installed, an additional softkey appears at the second
softkey from the left. For the HP 54655A, the label is HP-IB Menu. For the
HP 54656A, the label is RS-232 Menu.
3 Press the HP-IB Menu softkey or the RS-232 Menu softkey.
For HP 54655A, a softkey menu with five choices appears. Four of them are
related to the configuration of the HP-IB interface.
•
Connect to Selects the device connected to the oscilloscope.
Selections include HP plotter, HP printer, and computer.
•
Factors Selects whether the printer or plotter output contains the
information concerning the basic settings of the oscilloscope.
•
Resolution Selects the resolution of the printer or plotter output.
•
Address Selects the HP-IB address of the instrument. Valid addresses
range from 0 to 30. To change the address, press the softkey or turn the
knob closest to the Cursors key.
•
Previous Menu Returns the oscilloscope to the previous softkey menu.
For HP 54656A, a softkey menu with six choices appears. Five of them are
related to the configuration of the RS-232 interface.
•
Connect to Selects the device that the oscilloscope is connected to.
Selections include HP plotter, HP printer, Epson compatible printer, and
computer.
•
Factors Selects whether the printer or plotter output contains the
information concerning the basic settings of the oscilloscope.
•
Resolution Selects the resolution of the printer or plotter output.
•
Baud Rate Selects the baud rate of the RS-232 port. Valid baud rates
are 1200, 2400, 9600, and 19200.
•
Handshake Selects the handshaking method used in the RS-232. DTR
refers to the hardware handshaking using the DTR signal line. XON refers
to the software handshaking using the XON/XOFF protocol.
•
Previous Menu Returns the oscilloscope to the previous softkey menu.
Operating the Test Automation Module
To configure the interface
18