To assign a testing mask to a step
Before using a testing mask in a test step, the mask must be created and
stored in one of the forty mask memories.
1 Press Print/Utility twice.
2 Press the Sequence Admin softkey.
3 Press the Define Sequence softkey.
4 Press the Step softkey, or turn the knob closest to the Cursors key,
to select the step number for the testing mask number.
5 Press the Use Mask softkey.
6 Press the Use Mask softkey, or turn the knob closest to the Cursors
key, to select the mask number for testing.
Selecting the mask immediately assigns it to the selected step.
Mask Template Testing operating hint
When a mask template is used in a step, it compares the signal of the lowest
number of all active channels against the mask template. For example, if
channels 1, 3, and 4 are active, channel 1 is compared against the mask
template; if channels 2, 3, and 4 are active, channel 2 is compared against the
mask template.
Mask template testing cannot be used in conjunction with the delay sweep, roll
mode, or math functions.
To use mask template testing in the XY mode, refer to the section "To use mask
template testing in the XY mode" in this chapter.