When the failure condition of inside is selected, instead of displaying the two
limit lines on the screen, the failure regions are displayed as enclosed
polygons. Figure 7 shows an example of selecting inside as the failure
condition.
Example of selecting inside as the failure condition
The mask template for the inside failure condition is created and edited the
same way as mask templates used for normal testing. When it is edited, two
limit lines are displayed. Failure regions can be changed by editing the two
limit lines. Figure 8 shows the two limit lines of the mask template shown in
Figure 7.
Figure 7
Operating the Test Automation Module
To set up the mask template testing for "Fail when In" testing
42