Theory of Operation 8
16800 Series Portable Logic Analyzers Service Guide 177
The logic analyzer service carries the following general
responsibilities:
System manager for the local frame.
Communicates with local GUI application software.
Listens for connection requests across the LAN (remote
connect from other systems).
Initialization of all local frame hardware including modules.
This initialization includes loading of all FPGAs that are not
loaded directly by boot-EEPROMs when power is first
applied to the system.
Hardware monitoring.
Control over the "Touch Off" front panel button (if Option 103
is installed).
The logic analyzer service must be started and fully initialized
before the application software will be allowed to run. The logic
analyzer drivers are not required for the application software to
run – however the software will not be able to go online with the
local hardware until the drivers are installed/running. In
addition, other machines will not be able to remote connect to
the machine and use its hardware if the drivers are not installed
and running.
Because the service should always be running regardless of
whether any user is logged onto the system, the logic analyzer
should always be responsive to remote connection requests
from other machines. The Touch Off button should also always
function (if Option 103 is installed).
When the GUI application is launched by a logged on user, the
application (by default) will attempt to connect to local
hardware and go "online". If the service is not started, the
application should either wait for the service to start (newer
software revisions) or generate a message box indicating the
service needs to be started (older software revisions). If the
drivers are not loaded, the GUI application will start in offline
mode.
Once the GUI application is running, the user can go online with
local hardware or the hardware in any other Agilent
168x/169x/168xx/1690x logic analyzer running the same
version of the logic analyzer software. The remote machine
must be available on the network.