42 1100 Series Capillary LC System Manual
2 Optimizing Performance
5 Starting with a compressibility setting of 10 × 10
-6
/bar increase the value
in steps of 10. Re-zero the integrator as required. The compressibility
compensation setting that generates the smallest pressure ripple is the
optimum value for your solvent composition.
6 Repeat step 1 through step 5 for the B channel of your capillary pump.
Optimize your compressibility settings by using the values for various solvents
listed in the following table:
.
Tabl e 6 Solvent Compressibility
Solvent (pure) Compressibility (10
-6
/bar)
Acetone 126
Acetonitrile 115
Benzene 95
Carbon tetrachloride 110
Chloroform 100
Cyclohexane 118
Ethanol 114
Ethyl acetate 104
Heptane 120
Hexane 150
Isobutanol 100
Isopropanol 100
Methanol 120
1-Propanol 100
To l u e n e 8 7
THF 95
Water 46
Komentarze do niniejszej Instrukcji