140 1100 Series Capillary LC System Manual
6Options
Installing the 0.1 - 2.5 ml/min flow capillary kit
1 Switch off the pump module.
2 Disconnect the capillary 01090-87308 between the mixer and the filter.
3 Connect the capillary G1375-87400 to the mixer outlet.
4 Connect the other end of the capillary G1375-87400 to the purge. valve
holder.
5 Install the purge valve holder on the pump head of channel A and fix it with
the screw.
6 Screw the purge valve assembly into the purge valve holder and locate the
outlet and waste.
7 Use the 14 mm wrench to tighten the purge valve assembly.
8 Remove the waste tube from the EMPV and install it to the waste outlet of
the purge valve.
9 Disconnect the capillary on the injection valve (port 1).
10 Connect the capillary G1375-87318 between the purge valve and the
injection valve (port 1).
11 Replace the capillary between the injection valve and the analytical head
with the capillary G1375-87312.
Frequency When the flow rate is higher than 200 µl/min
Tools re quired 4 mm open wrench (8710-1534)
Torque adapter G1315-45003
1/4 - 5/16 inch open wrench (8710-0510)
14 mm wrench (8710-1924)
Parts required Extended flow range kit (G1376-69707)
Purge valve assembly G1311-60009 (supplied in the accessory kit
G1376-68705)
Purge valve holder G1312-23200 (supplied in the accessory kit G1376-68705)
Purge valve holder screw 0515-0175 (supplied in the accessory kit
G1376-68705)
Washer 2190-0586 (supplied in the accessory kit G1376-68705)