172 1100 Series Capillary LC System Manual
7 Performance Specifications
Performance Specifications Agilent 1100 Series DAD
Reference conditions for data of Table 92:
• cell path length 10 mm, response time 2 s,
• flow 1 ml/min LC-grade Methanol,
• slit width 4 nm.
Linearity measured with caffeine at 265 nm.
Tabl e 92 Performance Specifications Agilent 1100 Series Diode Array Detector
Type Specification Comments
Detection type 1024-element photodiode array
Light source Deuterium and tungsten lamps
Wavelength range 190 – 950 nm
Short term noise
(ASTM)
*
Single and
Multi-Wavelength
Typically ±3×10
-5
AU at 254 nm at flow
rates <100 µl/min
For the 500 nl flow cell the
noise is 2-3 times higher
than with standard flow
cell
Drift 2 × 10
-3
AU/hr at 254 nm
Linear absorbance
range
> 2 AU (upper limit)
Wavelength accuracy ± 1 nm Self-calibration with
deuterium lines,
verification with holmium
oxide filter
Wavelength bunching 1 – 400 nm Programmable in steps of
1 nm
Slit width 1, 2, 4 , 8, 16 nm Programmable slit
Diode width < 1 nm
Flow cell 500 nanoliter: 0.5 µl volume, 10 mm cell path
length and 50 bar (725 psi) pressure
maximum