1100 Series Capillary LC System Manual 11
Installing your Capillary LC System 1
Install the Capillary Pump (G1376A)
1 Ensure the line power switch at the front of the capillary pump is off.
2 Place the capillary pump on top of the micro WPS. Make sure that the two
modules are interlocked correctly.
3 Connect the power cable to the power socket at the rear of the capillary
pump. Do not connect the power cable to power until you have finished the
hardware installation of all modules in the stack.
4 Connect the free end of the CAN-bus cable from the micro sampler to one of
the CAN-bus connectors at the rear of the capillary pump.
5 Connect the pre-terminated end of the pump-to-sampler capillary
(G1375-87310) to the flow sensor outlet of the capillary pump. Connect the
other end of this capillary to port 1 of the sampler injection valve.
6 Connect the 1/8 inch plastic EMPV waste tube to the barbed waste fitting of
the EMPV. Route the waste tube to an appropriate waste container.
WARNING
Connect the power cable to power only after you have finished the hardware
installation of all modules in the stack.
NOTE
Carefully route all capillaries so that they are not crushed or broken by module front covers.
Avoid excessive bending. Refer to Chapter 2 for advice on handling capillaries.