1100 Series Capillary LC System Manual 161
Options 6
Replacing or Cleaning Parts
CAUTION
The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet
windows at the quartz block.
1 Disconnect the capillaries from the capillary holder
and remove the flow cell.
2 Unscrew the cell body from the holder.
3 Unscrew the capillaries from the flow cell. DO NOT
use the adapter at this time!
4 Using a toothpick, press on the plastic part and slide
the quartz body out of the cell housing.