1100 Series Capillary LC System Manual 15
Installing your Capillary LC System 1
Install the Thermostat for 1100 Samplers (G1330B)
1 Place the thermostat for 1100 samplers (thermostat module) on the bench.
The thermostat module should be no more than 25cm (9.8 inches) from the
front edge of the bench. The thermostat module should be positioned as the
bottom module in the right-hand stack.
2 Connect one end of the thermostat-to-sampler cable (G1330-81600) to the
26-pin connector at the rear of the thermostat module.
3 Route the large-bore plastic corrugated condensation drain tube from the
front of the thermostat module directly into an appropriate waste
container.
.
CAUTION
Connect the power cable to the thermostat module power socket only after you have
connected the thermostat-to-sampler cable (G1330-81600) between the thermostat
module and the micro sampler. Failure to follow this warning will result in damage to
the electronics of the thermostat module and the sampler.
NOTE
It is important that the condensation drain tube provides a straight, unblocked path for the
condensation to drain. The tube should never be coiled. The tube should never be below the
level of liquid in the waste container