36 1100 Series Capillary LC System Manual
2 Optimizing Performance
Hints for the Micro Vacuum Degasser
If you are using the vacuum degasser for the first time, if the vacuum degasser
was switched off for any length of time (for example, overnight), or if the
vacuum degasser lines are empty, you should prime the vacuum degasser
before running an analysis.
The vacuum degasser can be primed by pumping solvent with the capillary
pump at high flow rate (2.5 ml/min). Priming the degasser is recommended,
when:
• vacuum degasser is used for the first time, or vacuum chambers are empty.
• changing to solvent that are immiscible with the solvent currently in the
vacuum chambers.
• capillary pump was turned OFF for a length of time (for example during
night) and volatile solvent mixtures are used.
For more information see the Reference Manual for the Agilent 1100 series
micro vacuum degasser.