Agilent Technologies N4962A Podręcznik Użytkownika Strona 74

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 88
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 73
Remote GPIB Interface
74
Serial BERT 12.5 Gb/s User Guide
Command
:SENSe:ROSCillator:PHASe {AUTO | {0 <= value <= 360}}
Type
Command
Description
Sets the error detector clock phase for sampling point. If AUTO, the N4962A
will automatically determine the optimum phase for best sampling point once,
and set the phase to that value. For data rates less than 5 GHz, the phase
detector is limited to 0, 90, 180, or 270 degree increments.
Parameters
{AUTO | {0 <= value <= 358}}
Example
OUTPUT 725; ":SENS:ROSC:PHAS AUTO"
OUTPUT 725; ":SENS:ROSC:PHAS?"
ENTER 725; value$
PRINT value$ ! should be a number 0-360
Command
:SENSe:SWEep:TIME?
Type
Query
Description
Returns the current measurement time setting; how long the measurement is
intended to run for, in seconds.
Results
{0.001 <= value <= 4194303.999}
Example
See below
Przeglądanie stron 73
1 2 ... 69 70 71 72 73 74 75 76 77 78 79 ... 87 88

Komentarze do niniejszej Instrukcji

Brak uwag