Agilent 35670A
Dynamic Signal Analyzer
Versatile two- or four-channel high-per for mance
FFT-based spectrum/network an a lyz er
122 µHz to 102.4 kHz 16-bit ADC
Data Sheet
Key Specifi cations
Frequency range 102.4 kHz 1 channel
51.2 kHz 2 channel
25.6 kHz 4 channel
Dynamic range 90 dB typical
Accuracy ±0.15 dB
Channel match ±0.04 dB and ±0.5 degrees
Real-time bandwidth 25.6 kHz/1 channel
Resolution 100, 200, 400, 800 & 1600 lines
Time capture > 6 Msamples
Source types Random, burst random, periodic chirp, burst chirp,
pink noise, sine, swept-sine ( Option1D2), arbitrary
( Option 1D4)
Przeglądanie stron 0
1 2 3 4 5 6 7 8 9 10

Podsumowanie treści

Strona 1 - Dynamic Signal Analyzer

Agilent 35670ADynamic Signal Analyzer Versatile two- or four-channel high-per for mance FFT-based spectrum/network an a lyz er122 µHz to 102.4 kHz 16-

Strona 2

Remove all doubtOur repair and calibration services will get your equip-ment back to you, performing like new, when promised. You will get full value

Strona 3

2Summary of Features on Standard InstrumentThe following features are standard with the Agilent 35670A:Instrument modesFFT analysis Histogram/timeCor

Strona 4 - Agilent 35670A Specifi cations

3Data storage functionsBuilt-in 3.5 in., 1.44-Mbyte flexible disk also supports 720-KByte disks, and 2 Mbyte NVRAM disk. Both MS-DOS® and HP-LIF forma

Strona 5 - Frequency alias responses

4Agilent 35670A Specifi cationsInstrument specifications apply after 15 minutes warm-up and within 2 hours of the last self-calibration. When the inter

Strona 6

5Full span FFT noise floor (typical)Flat top window, 64 RMS averages, 800 line display.Harmonic distortion <-80 dBfsSingle Tone (in band), ≤ 0 dBf

Strona 7 - +

6Single channel phasePhase accuracy relative to external trigger± 4.0 deg16 time averages center of bin, DC coupled 0 dBfs to -50 dBfs only0 Hz < f

Strona 8 - ( —————————————— ) ≤

7TriggerTrigger modes Internal, source, external (analog setting) GPIBMaximum trigger delayPost trigger 8191 secondsPre trigger 8191 sample periodsNo

Strona 9 - General Specifi cations

8Computed order tracking – Option 1D0 Maximum order x Maximum RPM( —————————————— ) ≤ 60Online (real time) 1 channel m

Strona 10 - Remove all doubt

9General specifi cationsSafety standards CSA certified for electronic test and measurement equip-ment per CSA C22.2, NO. 231 This product is designed f