Operation
Serial BERT 12.5 Gb/s User Guide 41
Text
Type
Description
Default
JitterI port.
Synth 1
o options: 0, 1
Internal clock setting: 0 = use external clock
(internal clock
turned off); 1 = use internal clock.
Err00000
Five
-digit error detector error counter, included
for low BER applications, or for testing external
error injection. The counter accumulates errors
since the detector was reset, and wraps around
99999. This
counter should not be used as the
authoritative number of errors counted.
ErAd OFF
o options: OFF, 1E0, 1E1, 1E2, …, 1E7
Digital error injection setting, which indicates
how many bit inversions occur per second. OFF
indicates no bit errors are injected, 1E
0 indicates
1 error per second, 1E7 indicates 10 million errors
per second. To find the injected BER, divide by
the clock frequency. For example, at a setting of
1E2 (100) errors per second, the BER will be 1E
-8
at a clock rate of 10 GHz, or 8E
-9 at 12.5 G
Hz. See
Table
14.
When Display
->Ø is selected from the display
selector, pressing the Adjust
->Config State PLUS
button will prompt the N4962A to auto
matically
determine the optimum detector clock phase.
This is detailed further in Section
4.6.
NoData 0
o options: 0, 1
NoData option helps with the initialization