Getting Started
14
Serial BERT 12.5 Gb/s User Guide
1.3.5 Performance Recommendations
When using differential-mode connections, ensure the cables are phase
balanced
Differential connectors may be used single-ended if unused ports are
terminated in 50 Ω
Use high quality cables and connector savers (or adaptors)
Keep cable lengths short and minimize number of cable bends
Use a 7-10 in-lbs torque wrench when attaching connectors
Terminate all unused RF connectors with 50 Ω terminations
1.4 Connect the Hardware
Connect power cord to adaptor to N4962A; plug in the power cord
Connect TX CKO to TX CKI with included coaxial cable
Connect RX CKO to RX CKI with included coaxial cable
Connect the OUT (and optionally OUT¯¯¯ to your DUT input
Connect your DUT output to IN (and optionally IN¯¯)
1.5 Turn It On
Turn on the power switch from the back panel
Press the Data Path > Select Path button and select Enabled
o This will turn on the PRBS output; when Disabled is selected, the
instrument is in a 'safe mode' and no output is generated
Press the Display > Scroll button and select Ampl (V)
Press the Adjust > Ampl +/- buttons to set the single-ended output
voltage