Operation
Serial BERT 12.5 Gb/s User Guide 51
4.5.4 Low BER Test
To verify that the error detector can correctly count a small number of errors,
use the PRBS error injection capability to insert a known number of errors into
the bit stream, as follows:
Put the system in self test mode, described in Section 4.5.1.
Press the Display > Scroll button and select ErAd OFF
Press the Adjust > Config State + button and select ErAd 1E0
Press the Receiver > On button and ensure the light is on
Observe the Error ε light (should blink once per second)
Press the Display > Scroll ↑ button and select Err (should show 1 per
second)
Press the Display > Scroll ↓ button and select BER (should show
1E-10)
Press the Display > Scroll button and select ε time (sec) (should be
increasing)
4.6 BER Detector Clock Phase Adjustment
The error detector clock phase can be electronically controlled to adjust the
detector sampling point; the periodic point at which the detector interprets the
incoming bit stream. Due to the unknown delay associated with cables and the
DUT in the data path, the incoming bit stream may not be sampled at the
optimal point.
The detector must be adjusted to find the optimal sampling point; an incorrect
sampling point will yield an artificially-high BER measurement, shown in Figure
14.
Figure 14. Sampling uncertainty due to clock phase misadjustment
IN, notIN
RX CKI
good
IN, notIN
RX CKI
uncertain