Agilent Technologies Ininiium 9000 Dokumentacja Strona 92

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 91
Chapter 5: Troubleshooting
How to Use the Power Fault LEDs
92
How to Use the Power Fault LEDs
The previous page showed where you should look through the rear fan slots to see the upper right hand
corner of the acquisition board. There are a series of 13 LEDs. They are separated into three sections of
four LEDs each and then the Over Temperature LED is by itself at the bottom (as shown in Figure 5-5).
Figure 5-5
By looking at the pattern of LEDs illuminated after a power failure, you can determine the cause of the
fault. However, in most cases, only assembly level repair is required which greatly simplfies decoding
these LEDs. An in-depth discussion of this LEDs will first be presented so you understand how they
operate and then the assembly level repair procedures will be discussed.
The LEDs must be interpreted in groups of four (as shown in the layout above) with the upper two LEDs
Przeglądanie stron 91
1 2 ... 87 88 89 90 91 92 93 94 95 96 97 ... 165 166

Komentarze do niniejszej Instrukcji

Brak uwag