Chapter 4: Testing Performance
Performing Self-Test and Calibration
49
Performing Self-Test and Calibration
1 Warm up the oscilloscope for at least 30 minutes
2 Perform self tests
a Pull down the Utilities menu and select Self Test.
b Select Scope Self Test from the Self Test list.
c Click on Start Self Test to start the self test procedure.
If any of the self-tests fail, ensure that the failure is diagnosed and repaired before calibrating
and testing performance.
3
Perform calibration. See “To run the self calibration” on page 45.