Chapter 4: Testing Performance
Offset Accuracy Test
52
Offset Accuracy Test
CAUTION Ensure that the input voltage to the oscilloscope never exceeds ±5 V.
Specifications
Equipment Required
The offset accuracy specification consists of two terms: offset gain and zero error. The offset gain
characteristic is ±1.25% of channel offset and the zero error characteristic is ±1% of full scale. The
offset accuracy test procedure tests the zero error and offset accuracy.
Procedure
Zero Error Test 1 Disconnect all cables from the scope channel inputs.
2 Press Default Setup, then configure the scope as follows:
a Pull down the Setup menu and select Acquisition.
Let the oscilloscope warm up before testing
The oscilloscope and other equipment under test must be warmed up (with the oscilloscope
application running) for at least 30 minutes prior to the start of any performance test.
Offset Accuracy ±(1.25% of channel offset + 1% of full scale + 1 mV)
50 Ohm Input Impedance: Full scale is defined as 8 vertical divisions. Magnification is
used below 10 mV/div. Below 10 mV full scale is defined as 80 mV. The major scale set-
tings are 5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV and 1 V.
1 MOhm Input Impedance: Full scale is defined as 8 vertical divisions. The major scale
settings are 5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, and 5 V.
Description Critical Specifications Recommended
Model/
Part Numbers
Digital Multimeter DC voltage measurement accuracy better than ±0.1% of
reading
Agilent 34401A or
Agilent 3458A
Cable Assembly
(2 required)
50 characteristic impedance, BNC (m) connectors Agilent 8120-1840
DC Power Supply 100V Agilent 6614C
Adapter BNC Tee (m)(f)(f) Agilent 1250-0781
2 Adapters 2 BNC (f) to dual banana Agilent 1251-2277