Chapter 4: Testing Performance
Offset Accuracy Test
57
16
Set the channel 1 offset value to 60.0 mV and the impedance to 50 W (or to an offset of 2 V and
the impedance to 1 M
W if this is your second pass through this test and you are using the 1 MW input
impedance version)
by:
a Pull down the Setup menu and select Channel 1 or click the Channel 1 setup icon.
b Click the Offset control arrows to change the offset value or click on the offset value and enter
60.0 mV in the dialog box (or 2 V for the 1 M
W impedance version).
c Select the 50
W radio button in the Impedance section of the Channel 1 Setup dialog box.
17
Set the DC supply voltage (External Supply Setting) to +60.0 mV (or +2 V for the 1 MW
version of this test).
18 Press the Clear Display key on the scope, wait for the #Avgs value (top left corner of screen)
to return to 256 and then record the DMM voltage reading as V
DMM+
and the scope Vavg
reading as V
Scope+
in the Offset Accuracy Test section of the Performance Test Record (either
the 50 W or the 1 MW version depending on which version of the test you are on). Fill in the
V
error+
column by using the following equation:
V
error+
= V
DMM+
- V
scope+
.
V
error+
must be within the limits specified by the corresponding Offset Accuracy Limit listed
in the table.