Chapter 4: Testing Performance
Analog Bandwidth - Maximum Frequency Check
65
Analog Bandwidth - Maximum Frequency Check
CAUTION Ensure that the input voltage to the oscilloscope never exceeds ±5 V.
Specification
Equipment Required
Let the oscilloscope warm up before testing
The oscilloscope under test and other test equipment must be warmed up (with the oscilloscope
application running) for at least 30 minutes prior to the start of any performance test.
Analog Bandwidth (-3 dB)
DSO/MSO9404A 4 GHz
DSO/MSO9254A 2.5 GHz
DSO/MSO9104A 1 GHz
DSO/MSO9064A 600 MHz
Description Critical Specifications Recommended Model/
Part Numbers
Microwave CW
Generator
Maximum Frequency  GHz
Power range: -20 dBm to +16 dBm into 50
Output resistance = 50
Agilent E8257D with Opt 520
Power Splitter 2 Resistor Power Splitter (f)
Max Frequency 18 GHz
Agilent 11667B
Power Meter Agilent E-series with power sensor compatibility Agilent E4418B or E4419B
Power Sensor Maximum Frequency 14 GHz
Power range: -24 dBm to +16 dBm
Agilent E4413A
Microwave Cable 50 Characteristic Impedance
3.5 mm (m) to 3.5 mm (m) SMA connectors
Max Frequency 18 GHz
Agilent 8120-4948
N-Type to SMA
Adapter
N-type to 3.5 mm (m) SMA Agilent 1250-1750
SMA to BNC
Adapter
SMA (m) to Precision BNC (No Substitute) Agilent 1250-1200