Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 627
:TRIGger:USB:TRIGger
(see page 798)
Command Syntax
:TRIGger:USB:TRIGger <value>
<value> ::= {SOP | EOP | ENTersuspend | EXITsuspend | RESet}
The :TRIGger:USB:TRIGger command sets where the USB trigger will occur:
SOP — Start of packet.
EOP — End of packet.
ENTersuspend — Enter suspend state.
EXITsuspend — Exit suspend state.
RESet — Reset complete.
Query Syntax
:TRIGger:USB:TRIGger?
The :TRIGger:USB:TRIGger? query returns the current USB trigger value.
Return Format
<value><NL>
<value> ::= {SOP | EOP | ENTersuspend | EXITsuspend | RESet}
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:MODE" on page 485
":TRIGger:USB:SPEed" on page 626