Agilent Technologies InfiniiVision 6000 Series Instrukcja Obsługi Strona 542

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 934
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 541
542 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:I2S:CLOCk:SLOPe
(see page 798)
Command Syntax
:TRIGger:I2S:CLOCk:SLOPe <slope>
<slope> ::= {NEGative | POSitive}
The :TRIGger:I2S:CLOCk:SLOPe command specifies which edge of the I2S
serial clock signal clocks in data.
NEGative — Falling edge.
POSitive — Rising edge.
Query Syntax
:TRIGger:I2S:CLOCk:SLOPe?
The :TRIGger:I2S:CLOCk:SLOPe? query returns the current I2S clock slope
setting.
Return Format
<slope><NL>
<slope> ::= {NEG | POS}
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:I2S:ALIGnment" on page 540
":TRIGger:I2S:RWIDth" on page 548
":TRIGger:I2S:TWIDth" on page 554
":TRIGger:I2S:WSLow" on page 555
Przeglądanie stron 541
1 2 ... 537 538 539 540 541 542 543 544 545 546 547 ... 933 934

Komentarze do niniejszej Instrukcji

Brak uwag