Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 407
:POD<n>:THReshold
(see page 798)
Command Syntax
:POD<n>:THReshold <type>[<suffix>]
<n> ::= An integer, 1 or 2, is attached as a suffix to the command and
defines the group of channels that are affected by the command.
<type> ::= {CMOS | ECL | TTL | <user defined value>}
<user defined value> ::= -8.00 to +8.00 in NR3 format
<suffix> ::= {V | mV | uV}
POD1 ::= D0-D7
POD2 ::= D8-D15
TTL ::= 1.4V
CMOS ::= 2.5V
ECL ::= -1.3V
The :POD<n>:THReshold command sets the threshold for the specified
group of channels. The threshold is used for triggering purposes and for
displaying the digital data as high (above the threshold) or low (below the
threshold).
Query Syntax
:POD<n>:THReshold?
The :POD<n>:THReshold? query returns the threshold value for the
specified group of channels.
Return Format
<threshold><NL>
<threshold> ::= Floating point number in NR3 format
See Also "Introduction to :POD<n> Commands" on page 404
":DIGital<n>:THReshold" on page 255
":TRIGger[:EDGE]:LEVel" on page 513
Example Code
' THRESHOLD - This command is used to set the voltage threshold for
' the waveforms. There are three preset values (TTL, CMOS, and ECL)
' and you can also set a user-defined threshold value between
' -8.0 volts and +8.0 volts.
'
' In this example, we set channels 0-7 to CMOS, then set channels
' 8-15 to a user-defined 2.0 volts, and then set the external trigger
' to TTL. Of course, you only need to set the thresholds for the
' channels you will be using in your program.