Przeglądanie stron 488
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 489
:TRIGger:SWEep
(see page 798)
Command Syntax
:TRIGger:SWEep <sweep>
<sweep> ::= {AUTO | NORMal}
The :TRIGger:SWEep command selects the trigger sweep mode.
When AUTO sweep mode is selected, a baseline is displayed in the absence
of a signal. If a signal is present but the oscilloscope is not triggered, the
unsynchronized signal is displayed instead of a baseline.
When NORMal sweep mode is selected and no trigger is present, the
instrument does not sweep, and the data acquired on the previous trigger
remains on the screen.
Query Syntax
:TRIGger:SWEep?
The :TRIGger:SWEep? query returns the current trigger sweep mode.
Return Format
<sweep><NL>
<sweep> ::= current trigger sweep mode
See Also "Introduction to :TRIGger Commands" on page 479
NOTE
This feature is called "Mode" on the instrument's front panel.