Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 519
:TRIGger:FLEXray:BAUDrate
(see page 798)
Command Syntax
:TRIGger:FLEXray:BAUDrate <baudrate>
<baudrate> ::= {2500000 | 5000000 | 10000000}
The :TRIGger:FLEXray:BAUDrate command specifies the baud rate as
2.5Mb/s, 5Mb/s, or 10Mb/s.
Query Syntax
:TRIGger:FLEXray:BAUDrate?
The :TRIGger:FLEXray:BAUDrate? query returns the current baud rate
setting.
Return Format
<baudrate><NL>
<baudrate> ::= {2500000 | 5000000 | 10000000}
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:FLEXray Commands" on page 517
NOTE
This command is only valid on 4 (analog) channel oscilloscope models when the FlexRay
triggering and serial decode option (Option FLX) has been licensed.