Przeglądanie stron 570
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 571
:TRIGger:LIN:PATTern:FORMat
(see page 798)
Command Syntax
:TRIGger:LIN:PATTern:FORMat <base>
<base> ::= {BINary | HEX | DECimal}
The :TRIGger:LIN:PATTern:FORMat command sets the entry (and query)
number base used by the :TRIGger:LIN:PATTern:DATA command. The
default <base> is DECimal.
Query Syntax
:TRIGger:LIN:PATTern:FORMat?
The :TRIGger:LIN:PATTern:FORMat? query returns the currently set
number base for LIN pattern data.
Return Format
<base><NL>
<base> ::= {BIN | HEX | DEC}
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:LIN:PATTern:DATA" on page 568
":TRIGger:LIN:PATTern:DATA:LENGth" on page 570
NOTE
This command is only valid when the automotive CAN and LIN serial decode option (Option
AMS) has been licensed.