462 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:SYSTem:PRECision
(see page 798)
Command Syntax
:SYSTem:PRECision <value>
<value> ::= {{1 | ON} | {0 | OFF}}
The :SYSTem:PRECision command turns the oscilloscope's precision
analysis setting on or off.
OFF (0) — provides the maximum oscilloscope waveform update rate by
performing measurements and math functions on a 1000- point
measurement record.
ON (1) at the expense of oscilloscope waveform update rate, this
setting allows measurements and math functions to be performed on a
precision analysis record (see ":WAVeform:POINts:MODE" on page 643).
The precision analysis setting is OFF after a *RST command.
Precision analysis is not available when:
Realtime sampling mode is off.
Averaging or High Resolution acquisition modes are selected.
XY or Roll time modes are selected.
Query Syntax
:SYSTem:PRECision?
The :SYSTem:PRECision? query returns the current precision analysis
setting.
Return Format
<value><NL>
<value> ::= {1 | 0}
See Also "Introduction to :SYSTem Commands" on page 457
":WAVeform:POINts:MODE" on page 643
"*RST (Reset)" on page 140