Agilent Technologies InfiniiVision 6000 Series Instrukcja Obsługi Strona 570

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 934
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 569
570 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:LIN:PATTern:DATA:LENGth
(see page 798)
Command Syntax
:TRIGger:LIN:PATTern:DATA:LENGth <length>
<length> ::= integer from 1 to 8 in NR1 format
The :TRIGger:LIN:PATTern:DATA:LENGth command sets the number of
8- bit bytes in the LIN data string. The number of bytes in the string can
be anywhere from 0 bytes to 8 bytes (64 bits). The value for these bytes is
set by the :TRIGger:LIN:PATTern:DATA command.
Query Syntax
:TRIGger:LIN:PATTern:DATA:LENGth?
The :TRIGger:LIN:PATTern:DATA:LENGth? query returns the current LIN
data pattern length setting.
Return Format
<count><NL>
<count> ::= integer from 1 to 8 in NR1 format
Errors "- 241, Hardware missing" on page 757
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:LIN:PATTern:DATA" on page 568
":TRIGger:LIN:SOURce" on page 574
NOTE
This command is only valid when the automotive CAN and LIN serial decode option (Option
AMS) has been licensed.
Przeglądanie stron 569
1 2 ... 565 566 567 568 569 570 571 572 573 574 575 ... 933 934

Komentarze do niniejszej Instrukcji

Brak uwag