Chapter 3 63
Operation
Using an 71000 Series Spectrum Analyzer with an 70907B, 70909A, or 70910A
Operation
Using an 71000 Series Spectrum Analyzer
with an 70907B, 70909A, or 70910A
Preliminary Operation
NOTE This section provides information for the 70907B, however the operation of
the 70909A and 70910A is similar. Minor differences may be noted, but the
necessary deviations from these exact instructions will be obvious.
Compatibility Reference
Refer to Chapter 2, “Installation,” on page 27 for information relating to 71000 Series
Modular Spectrum Analyzer System compatibility with the 11974 Series.
Equipment Connection
Connect the equipment as shown in Figure 3-6. and use the following checklist to verify
the connections.
• 11974 LO IN to 70907B LO OUTPUT (SMA cable, 5061-5458)
• 11974 IF OUT to 70907B IF INPUT (SMA cable, 5061-5458)
• 11974 TUNE IN to 70907B PRESEL TUNE/SPAN (BNC cable, 8120-1839)
• 11974 POWER SUPPLY to 11974-60028
Figure 3-6. 11974 Connections to the 71000 Series Spectrum Analyzer