Chapter 3 61
Operation
Using an 8566B Spectrum Analyzer
Operation
Operation
Band Selection
1. If necessary, mate the 11974 to the spectrum analyzer being used for preselected
external mixing by using the Frequency Tracking Calibration procedure above.
2. Set the spectrum analyzer to the appropriate millimeter band by using the key
sequence listed below:
for an 11974A:
SHIFT
for an 11974Q: SHIFT
for an 11974U: SHIFT
for an 11974V: SHIFT
NOTE The above band selection key sequence must be used to enter the external
mixing band of operation. Do not enter in the center frequency of the band
directly.
Amplitude Calibration
The 8566B uses a single-point conversion-loss factor per band. The conversion-loss number
is saved with an instrument preset, when the band is exited, and with the SAVE and
RECALL functions.
Enter the conversion loss at the desired frequency from either the calibration label on the
bottom of the 11974 or the supplied calibration sheet using the following key sequences.
1.
SHIFT, 5, and MHz
2. Enter the conversion loss value of the desired frequency, then terminate the entry with
dB.
Preselector Peaking
1. Place the active marker on the desired signal to be peaked using
PEAK SEARCH or
MARKER
NORMAL.
TIP To complete the preselector-peaking routine more quickly, place the signal at
center screen and reduce the frequency span until the sweep time is
approximately 200 ms.
2. Peak the desired signal by pressing
PRESEL PEAK.