40 Chapter 2
Installation
Electrostatic Discharge Information
Installation
Reducing ESD Damage
The following are suggestions that may help reduce the amount of ESD damage that can
occur during testing and servicing instruments.
PC Board Assemblies and Electronic Components
Handle these items at a static-safe workstation.
Store or transport these items in static-shielding containers.
CAUTION Do not use erasers to clean the edge-connector contacts. Erasers generate
static electricity and degrade the electrical quality of the contacts by
removing the thin gold plating.
Do not use paper of any kind to clean the edge-connector contacts. Paper or
lint particles left on the contact surface can cause intermittent electrical
connections.
Do not touch the edge-connector contacts or trace surfaces with bare hands.
Always handle board assemblies by the edges.
PC board assembly edge-connector contacts may be cleaned by using a lint-free cloth with
a solution of 80 percent electronics-grade isopropyl alcohol and 20 percent deionized water.
This procedure should be performed at a static-safe workstation.
WARNING Isopropyl alcohol is extremely flammable, causes irritation, may
cause eye damage, and is harmful if swallowed or inhaled. It may be
harmful if absorbed through the skin. Keep away from heat, sparks,
and flame. Avoid contact with eyes, skin, clothing. Avoid breathing
vapor. Keep in tightly closed container. Use with adequate
ventilation. Wash thoroughly after handling. In case of fire, use
alcohol foam, dry chemical, or carbon dioxide: water may be
ineffective. In case of spill, soak up with sand or earth. Flush spill
area with water.
Disposal Instructions: Dispose in accordance with all applicable
federal, state, and local environmental regulations.
Test Equipment
Before connecting any coaxial cable to an instrument connector for the first time each
day, momentarily short together the center and outer conductors of the cable.
Before touching the center pin of any connector and before removing any assembly from
the instrument, ensure proper use of a grounded, resistor-isolated, wrist strap.
To prevent buildup of static charge, ensure that all instruments are properly
earth-grounded.