120 Chapter 4
Performance Tests
3. Multiple Response Rejection Test
Performance Tests
11.Set the spectrum analyzer to the appropriate millimeter band by using the following
key sequence:
for an 11974A:
SHIFT
for an 11974Q: SHIFT
for an 11974U: SHIFT
for an 11974V: SHIFT
12.Set the spectrum analyzer controls as follows:
SHIFT, 5, MHz, 0, and dB (sets external mixer conversion loss to 0 dB)
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 kHz
ENTER dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dB
13.Set the spectrum analyzer reference level by pressing
REFERENCE LEVEL and entering
the corresponding value for the band listed below:
for an 11974A:
40 dBm
for an 11974Q:
40 dBm
for an 11974U:
40 dBm
for an 11974V:
55 dBm
14.Set the spectrum analyzer center frequency by pressing
CENTER FREQUENCY and
entering the corresponding value for the band listed below:
for an 11974A: 26.5 GHz
for an 11974Q: 33 GHz
for an 11974U: 40 GHz
for an 11974V: 50 GHz
15.On the spectrum analyzer, press the following keys:
MARKER
NORMAL
PEAK SEARCH
PRESEL PEAK
16.When the preselector-peak routine is complete, press PEAK SEARCH.
17.Record the marker frequency in Table 4-9 on page 144 as Signal Frequency #1.
18.On the spectrum analyzer, press MARKER
.