576 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
Minimum Response –2.0 dB (17)________ ±0.59 dB
Peak-to-Peak Response (18)________ 3.6 dB
±0.59 dB
Option UKB
ac Coupled
Band 0
(10 MHz to 3.0 GHz)
Maximum Response (19)________ 0.5 dB
±0.11 dB
Minimum Response –0.5 dB (20)________
±0.11 dB
Peak-to-Peak Response (21)________ 1.0 dB
±0.11 dB
Option UKB
ac Coupled
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39440871 or greater)
Maximum Response (22)________ 0.5 dB
±0.11 dB
Minimum Response –0.5 dB (23)________
±0.11 dB
Peak-to-Peak Response (24)________ 1.0 dB
±0.11 dB
Option UKB
ac Coupled
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39440871 or greater)
Maximum Response (25)________ 0.5 dB
±0.11 dB
Minimum Response –0.5 dB (26)________ ±0.11 dB
Peak-to-Peak Response (27)________ 1.0 dB
±0.11 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty