Agilent Technologies E1401-90493 Instrukcja Użytkownika Strona 222

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 642
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 221
222 Chapter 2
Performance Verification Tests
26. Frequency Response (Preamp On): Agilent E4402B
calibration factor of the reference sensor at 1 MHz.
4. On the power meter, set the Channel B calibration factor to 100%.
Do not change this calibration factor during this test.
5. Connect the equipment as shown in Figure 2-38. Use the function
generator as the source. Note that the reference sensor connects to
the 20 dB fixed attenuator.
Figure 2-38 Source/Splitter Characterization Setup
6. Set the function generator frequency to 1 MHz and amplitude to
450 mV rms (approximately 6 dBm).
7. Adjust the source amplitude to obtain a Channel A power meter
reading of –20 dBm
±0.1 dB.
8. Record the Channel A and Channel B power meter readings in
Table 2-60.
9. Tune the source to the next frequency in Table 2-60.
10.On the power meter, set the Channel A calibration factor to the
calibration factor of the reference sensor for the current source
frequency.
11.Adjust the source amplitude to obtain a Channel A power meter
reading of –20 dBm
±0.1 dB.
12.Record the Channel A and Channel B power meter readings in
Table 2-60.
13.Repeat step 9 through step 12 for frequencies up through 10 MHz.
14.Replace the function generator with the synthesized sweeper.
Przeglądanie stron 221
1 2 ... 217 218 219 220 221 222 223 224 225 226 227 ... 641 642

Komentarze do niniejszej Instrukcji

Brak uwag