Chapter 2 101
Performance Verification Tests
13. Input Attenuation Switching Uncertainty
Figure 2-13 Input Attenuator Switching Uncertainty Test Setup
CAUTION Use only 75 cables, connectors, or adapters on instruments with 75
connectors, or the connectors will be damaged.
Procedure
Calculate Actual Attenuation Values
1. From the calibration data supplied with the 1 dB step attenuator,
enter into Column 2 of Table 2-18 the actual attenuation for the
corresponding nominal attenuation settings. If the calibration data
does not indicate an actual attenuation value for the 0 dB setting,
enter 0 dB.
NOTE The Agilent 8494G programmable attenuator has four attenuator
sections consisting of 1 dB, 2 dB, 4 dB, and 4 dB attenuators. If using
the Agilent 8494G programmable attenuator, enter the calibration data
for the section three 4 dB step rather than the section four 4 dB step.
2. From the calibration data supplied with the 10 dB step attenuator,
enter into Column 4 of Table 2-18 the actual attenuation for the
corresponding nominal attenuation settings. If the calibration data
does not indicate an actual attenuation value for the 0 dB setting,
enter 0 dB.
NOTE The Agilent 8496G programmable attenuator has four attenuator
sections consisting of 10 dB, 20 dB, 40 dB, and 40 dB attenuators. If