466 Chapter 3
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
Option UKB
ac Coupled
Band 0
(100 kHz to 3.0 GHz)
Maximum Response (10)________ 0.50 dB
±0.11 dB
Minimum Response –0.50 dB (11)________
±0.11 dB
Peak-to-Peak Response (12)________ 1.0 dB ±0.11 dB
Option UKB
ac Coupled
Band 0A
(800 MHz to 1.0 GHz)
Maximum Response (13)________ 0.50 dB
±0.11 dB
Minimum Response –0.50 dB (14)________
±0.11 dB
Peak-to-Peak Response (15)________ 1.0 dB
±0.11 dB
Option UKB
ac Coupled
Band 0B
(1.7 GHz to 2.0 GHz)
Maximum Response (16)________ 0.50 dB
±0.11 dB
Minimum Response –0.50 dB (17)________ ±0.11 dB
Peak-to-Peak Response (18)________ 1.0 dB
±0.11 dB
0 to 55
° C
Non-Option UKB
Band 0
(9 kHz to 3.0 GHz)
Maximum Response (1)________ 0.76 dB
±0.19 dB
Minimum Response –0.76 dB (2)________
±0.19 dB